Intercomparison of Dobson Ozone Spectrophotometer in Yonsei University, Korea
Koji
MIYAGAWA, Jhoon Kim*
and Hi Ku Cho*
*Yonsei University, Korea |
|
Abstract
At Yonsei University in Korea, the ozone
observation using Dobson ozone spectrophotometer is the 20th year since the
commencement in 1984. The international ozone workshop (Joint International
Workshop on Ozone and ILAS-II Science Team Meeting), which commemorated the 20th
anniversary of ozone observation in Korea, was held at Yonsei University in
November 2004. Associating with the workshop, Yonsei University performed the
instrument calibration of Dobson No.124 for maintaining the measurement accuracy
of ozone observation. The intercomparison of Dobson in Seoul was the first trial
performed by shipping the traveling reference Dobson calibrated by the Asian
regional standard No.116 in Japan. The Dobson No.124 is maintained at Global
Environment Laboratory, Yonsei University and the result of ozone observation is
reported to the World Ozone and Ultraviolet Radiation Data Centre (WOUDC). In
the Dobson intercomparison at Yonsei University, comparison measurements,
instrument maintenance, in-depth lamp tests, and the technical study were
carried out during the two weeks. Finally, the Dobson No.124 was adjusted to an
accuracy of less than 1 percent in total ozone comparing with the traveling
reference. As in this case in Korea, on-site Dobson calibration by traveling
reference instrument and the expert avoids possible characteristic shift of the
comparison instrument due to transportation. Additional merits are the
best-suited technical advice offered through the survey on the site environment,
personnel, operation practice and data handling, as well as the information
sharing of the latest technology among the participants. The practical
discussion and strategy in accordance with the needs of the site is the key to
successful station updating.
|